Portable Digital small Blood sample Glucose Meter
DM5279
INTERFERING SUBSTANCES
(1)Ibuprofen (no more than 50 mg/dL), salicylic acid (no more than 60 mg/dL), creatinine (no more than 5 mg/dL), uric acid (no more than 23.5 mg/dL), ascorbic acid (no more than 3 mg/dL), acetaminophen (no more than 20 mg/dL), dopamine (no more than 0.09 mg/dL), levodopa (no more than 3 mg/dL), xylose (no more than 5 mg/dL), galactose (no more than 100 mg/dL), maltose (no more than 200 mg/dL), heme (no more than 500 mg/dL), conjugated bilirubin (no more than 50 mg/dL), unconjugated bilirubin (no more than 40 mg/dL), triglyceride (no more than 3000 mg/dL), and cholesterol (no more than 500 mg/dL) have no significant interference with the test results, and if exceeding the ranges, they may lead to a large deviation in the blood glucose test results.
(2) Do not use anticoagulants containing sodium fluoride for blood glucose testing.
(3) Glutathione (reduced) and pralidoxime iodide at therapeutic concentrations may affect the blood glucose test results, so people taking these two drugs should not use this product for blood glucose testing.
(4) The range of hematocrit (HCT) required by this product is 15%–70%. If HCT is out of this range, please do not use this product for testing.